Publication: Repeated procedures at the generator pocket are a determinant of implantable cardioverter-defibrillator infection.
No Thumbnail Available
Identifiers
Date
2017-06-21
Authors
Arana-Rueda, Eduardo
Pedrote, Alonso
Frutos-López, Manuel
Acosta, Juan
Jauregui, Beatriz
García-Riesco, Lorena
Arce-León, Álvaro
Gómez-Pulido, Federico
Sánchez-Brotons, Juan A
Gutiérrez-Carretero, Encarnación
Advisors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Rates of cardiac-device infections have increased in recent years, but the current incidence and risk factors for infection in patients with implantable cardioverter-defibrillators (ICDs) are not well known. The increasing number of ICD infections is related to accumulated pocket manipulations over time. This single-center, prospective study included patients that underwent ICD implantation from 2008 to 2015. The endpoint was time to infection. Multivariate analysis was performed to identify independent risk factors related to infection. The study included a total of 570 patients, of whom 419 (73.5%) underwent a first implantation. Mean age was 59 ± 14 years, and 80% were male. During a median follow-up of 36 months (interquartile range, 18-61 months; 1887 patient-years), infection was identified in 26 patients (4.56%), an incidence of 14.9 × 1000 patient-years. Median time to infection was 9.7 months (interquartile range, 1.35-23.4 months), and 38.5% were late infections (beyond 12 months of follow-up). In patients with replacement implants, the incidence was 3-fold higher than in first implantations (27.7 vs 9.1 × 1000 patient-years; P = 0.002). Cox regression identified 2 independent predictors of ICD infection: cumulative number of interventions at the generator pocket (hazard ratio: 1.92, 95% confidence interval: 1.42-2.6, P The incidence of infection in ICD patients is greater than previously reported, largely due to late infections. Each new cumulative intervention at the same generator pocket nearly doubles the risk of infection.
Description
MeSH Terms
Adult
Aged
Defibrillators, Implantable
Electric Countershock
Female
Humans
Incidence
Kaplan-Meier Estimate
Male
Middle Aged
Multivariate Analysis
Proportional Hazards Models
Prospective Studies
Prosthesis-Related Infections
Registries
Risk Assessment
Risk Factors
Spain
Time Factors
Treatment Outcome
Aged
Defibrillators, Implantable
Electric Countershock
Female
Humans
Incidence
Kaplan-Meier Estimate
Male
Middle Aged
Multivariate Analysis
Proportional Hazards Models
Prospective Studies
Prosthesis-Related Infections
Registries
Risk Assessment
Risk Factors
Spain
Time Factors
Treatment Outcome
DeCS Terms
CIE Terms
Keywords
Defibrillation-ICD, Incidence, Infection Complication, Noninvasive Risk Assessment Tests