Publication: Realistic aortic phantom to study hemodynamics using MRI and cardiac catheterization in normal and aortic coarctation conditions.
No Thumbnail Available
Identifiers
Date
2016-08-13
Authors
Urbina, Jesus
Sotelo, Julio A
Springmüller, Daniel
Montalba, Cristian
Letelier, Karis
Tejos, Cristian
Irarrazaval, Pablo
Andia, Marcelo E
Razavi, Reza
Valverde, Israel
Advisors
Journal Title
Journal ISSN
Volume Title
Publisher
John Wiley & Sons, Inc.
Abstract
To design and characterize a magnetic resonance imaging (MRI)-compatible aortic phantom simulating normal and aortic coarctation (AoCo) conditions and to compare its hemodynamics with healthy volunteers and AoCo patients .Materials and MethodsThe phantom is composed of an MRI-compatible pump, control unit, aortic model, compliance chamber, nonreturn, and shutoff valves. The phantom without and with AoCo (13, 11, and 9 mm) was studied using 2D and 3D phase-contrast data and with a catheterization unit to measure pressures . The phantom data were compared with the mean values of 10 healthy volunteers and two AoCo patients .ResultsHemodynamic parameters in the normal phantom and healthy volunteers were: heart rate : 68/61 bpm, cardiac output : 3.5/4.5 L/min, peak flow and peak velocity (Vpeak) in the ascending aorta (AAo): 270/357 mL/s (significantly, P < 0.05) and 97/107 cm/s (not significantly, P = 0.16), and pressure in the AAo of the normal phantom of 131/58 mmHg. Hemodynamic parameters in the 13, 11, and 9 mm coarctation phantoms and Patients 1 and 2 were: heart rate : 75/75/75/97/78 bpm, cardiac output : 3.3/3.0/2.9/4.0/5.8 L/min, peak flow in the AAo: 245/265/215/244/376 mL/s, Vpeak in the AAo: 96/95/81/196/187 cm/s, Vpeak after the AoCo: 123/187/282/247/165 cm/s, pressure in the AAo: 124/56, 127/51, 133/50, 120/51 and 87/39 mmHg, and a trans-coarctation systolic pressure gradient: 7, 10, 30, 20, and 11 mmHg.ConclusionWe propose and characterize a normal and an AoCo phantom, whose hemodynamics , including velocity, flow, and pressure data are within the range of healthy volunteers and patients with AoCo. J. Magn. Reson. Imaging 2016;44:683–697.
Description
MeSH Terms
Aorta
Aortic Coarctation
Blood Flow Velocity
Cardiac Catheterization
Cardiac Imaging Techniques
Equipment Design
Equipment Failure Analysis
Humans
Magnetic Resonance Angiography
Phantoms, Imaging
Reproducibility of Results
Sensitivity and Specificity
Aortic Coarctation
Blood Flow Velocity
Cardiac Catheterization
Cardiac Imaging Techniques
Equipment Design
Equipment Failure Analysis
Humans
Magnetic Resonance Angiography
Phantoms, Imaging
Reproducibility of Results
Sensitivity and Specificity
DeCS Terms
Pacientes
Hemodinámica
Frecuencia cardíaca
Gasto cardíaco
Coartación aórtica
Presión sanguínea
Cateterismo
Hemodinámica
Frecuencia cardíaca
Gasto cardíaco
Coartación aórtica
Presión sanguínea
Cateterismo
CIE Terms
Keywords
3D PC-MRI, aortic coarctation, aortic phantom, catheterization, hemodynamic parameters
Citation
Urbina J, Sotelo JA, Springmüller D, Montalba C, Letelier K, Tejos C, et al. Realistic aortic phantom to study hemodynamics using MRI and cardiac catheterization in normal and aortic coarctation conditions. J Magn Reson Imaging. 2016 Sep;44(3):683-97.