Perez-Bartolome, FranciscoRocha-De-Lossada, CarlosSanchez-Gonzalez, Jose-MariaFeu-Basilio, SilviaTorras-Sanvicens, JosepPeraza-Nieves, Jorge2025-01-072025-01-072021-12-01https://hdl.handle.net/10668/26320This study examines agreement between the devices Anterion(R) and Pentacam HR(R) used for corneal and pupil measurements in healthy eyes. The parameters compared between the two devices were: anterior Km (D), anterior K2 (D), anterior K1 (D), anterior K1 axis (degrees), anterior astigmatism (D), anterior K max (D), posterior Km (D), posterior K2 (D), posterior K1 (D), posterior K1 axis (degrees), posterior astigmatism (D), CCT (mu m), thinnest point thickness (mu m), thinnest point X-coordinate (mm), thinnest point Y-coordinate (mm), pupil diameter (mm), pupil center-corneal vertex distance (mm) (angle kappa), pupil centroid angle (degrees), pupil centroid X-coordinate (mm), and pupil centroid Y-coordinate (mm). The Student's t test for independent samples identified significant differences (penAttribution 4.0 Internationalhttp://creativecommons.org/licenses/by/4.0/anterior-segment swept-sourceocular coherence tomographyScheimpflug imagingagreement analysispupil measurementsAngle kappaImplantationAberrationsThicknessLasikOctAnterior-Segment Swept-Source Ocular Coherence Tomography and Scheimpflug Imaging Agreement for Keratometry and Pupil Measurements in Healthy Eyesresearch article34945085open access10.3390/jcm102457892077-0383https://www.mdpi.com/2077-0383/10/24/5789/pdf?version=1639134657737630300001