RT Journal Article T1 Computation of the electron beam quality [Formula: see text] factors for the NE2571, NE2571A and NE2581A thimble ionization chambers using PENELOPE. A1 Erazo, F A1 Brualla, L A1 Lallena, A M K1 correction factors K1 Electron beams K1 NE2571 K1 NE2571A K1 NE2581A K1 PENELOPE AB The quality correction factor [Formula: see text] for electron beams was calculated for three thimble ionization chambers, namely, NE2571, NE2571A and NE2581A. The Monte Carlo code PENELOPE was used to estimate the overall correction factor fc,Q of these chambers for electron beams with nominal energies ranging between 6 and 22MeV, corresponding to a Varian Clinac 2100 C/D. A 60Co beam was used as reference quality Q0. Also eight monoenergetic electron beams reproducing the quality index R50 of the Clinac beams were considered. The [Formula: see text] factors were calculated as the ratio between fc,Q and [Formula: see text] . Those obtained for the NE2571 ionization chamber show a nice agreement with those calculated by Muir and Rogers with EGSnrc. As it occurred to other ionization chambers analyzed in previous works, the [Formula: see text] factors found for the monoenergetic beams are larger (smaller) than those corresponding to the Clinac beams at low (high) R50 values, the differences being slightly above 0.5%. Finally, the [Formula: see text] factors obtained in the case of the NE2571A chamber are systematically ∼0.5% below those of its predecessor chamber, the NE2571. YR 2017 FD 2017-05-16 LK http://hdl.handle.net/10668/11299 UL http://hdl.handle.net/10668/11299 LA en DS RISalud RD Apr 7, 2025