RT Journal Article T1 Influence of reduction in the elution times on HPLC glycohaemoglobin results A1 Leon-Justel, Antonio A1 Santotoribio, Jose D. A1 Dominguez-Pascual, Inmaculada A1 Delgado, Ana L. A1 Macias, Carlos A1 Herrera, Maria T. A1 Leal, Alfonso A1 Mangas, Miguel A. A1 Acosta, Domingo A1 Guerrero, J.M. K1 Cromatografía líquida de alta presión K1 Sangre AB ObjectivesTo compare HPLC methods with short and long elution times for HbA1c measurement in blood.MethodsComparison between G7-Tosoh (1.2 min); Bio-Rad-Variant-II-Turbo (1.3 min) and Arkray-HA-8160 (2.9 min).ResultsPassing–Bablok regression equations were: Y = 0.17 + 0.96X; Y = − 0.39 + 1.01X; Y = − 0.40 + 1.0X and the means of the differences using Bland–Altman Plot were 0.02; − 0.34; 0.32 for HA-8160/G7-Tosoh, HA-8160/Variant-II-Turbo and G7-Tosoh/Variant-II-Turbo, respectively.ConclusionsFaster elution methods had no problems on reproducibility of results obtained by slower elution methods. PB Elsevier SN 1873-2933 YR 2009 FD 2009-07-16 LK https://hdl.handle.net/10668/23207 UL https://hdl.handle.net/10668/23207 LA en NO Leon-Justel A, Santotoribio JD, Dominguez-Pascual I, Delgado AL, Macias C, Herrera MT, et al. Influence of reduction in the elution times on HPLC glycohaemoglobin results. Clin Biochem. 2009 Oct;42(15):1582-4. DS RISalud RD Apr 12, 2025