RT Journal Article T1 Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr. A1 Calvo, Jorge A1 Cano, María Eliecer A1 Pitart, Cristina A1 Marco, Francesc A1 Rodríguez-Martínez, José Manuel A1 Pascual, Alvaro A1 Martínez-Martínez, Luis K1 Escherichia coli K1 Aminoglicósidos K1 Enterobacteriaceae K1 Klebsiella pneumoniae K1 Errores diagnósticos K1 Antibacterianos AB The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted. PB American Society for Microbiology SN 0095-1137 YR 2011 FD 2011-09 LK http://hdl.handle.net/10668/1580 UL http://hdl.handle.net/10668/1580 LA en NO Calvo J, Cano ME, Pitart C, Marco F, Rodríguez-Martínez JM, Pascual A, et al. Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr. J Clin Microbiol. 2011; 49(9):3343-5 NO Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't; DS RISalud RD Apr 7, 2025