Publication:
Light-Dependent Resistors as Dosimetric Sensors in Radiotherapy.

dc.contributor.authorRomán-Raya, Juan
dc.contributor.authorRuiz-García, Isidoro
dc.contributor.authorEscobedo, Pablo
dc.contributor.authorPalma, Alberto J
dc.contributor.authorGuirado, Damián
dc.contributor.authorCarvajal, Miguel A
dc.date.accessioned2023-02-08T14:43:13Z
dc.date.available2023-02-08T14:43:13Z
dc.date.issued2020-03-11
dc.description.abstractSafe quality control of radiotherapy treatments lies in reliable dosimetric sensors. Currently, ionization chambers and solid-state diodes along with electrometers as readout systems are accomplishing this task. In this work, we present a well-known and low-cost semiconductor sensor, the light-dependent resistor (LDR), as an alternative to the existing sensing devices for dosimetry. To demonstrate this, a complete characterization of the response to radiation of commercial LDRs has been conducted in terms of sensitivity, reproducibility and thermal correction under different bias voltages. Irradiation sessions have been applied under the common conditions in radiotherapy treatments using a hospital linear accelerator. Moreover, the same electrometer used for the ionization chamber has also been successfully used for LDRs. In comparison with the sensitivity achieved for the ionization chamber (0.2 nC/cGy at 400 V bias voltage), higher sensitivities have been measured for the proposed LDRs, ranging from 0.24 to 1.04 nC/cGy at bias voltages from 30 to 150 V, with a reproducibility uncertainty among samples of around 10%. In addition, LDR temperature dependence has been properly modeled using the simple thermistor model so that an easy thermal drift correction of dose measurements can be applied. Therefore, experimental results show that LDRs can be a reliable alternative to dosimetric sensors with the advantages of low size, affordable cost and the fact that it could be adopted with minimal changes in routine dosimetry quality control since the same readout system is fully compatible.
dc.identifier.doi10.3390/s20061568
dc.identifier.essn1424-8220
dc.identifier.pmcPMC7146615
dc.identifier.pmid32168978
dc.identifier.pubmedURLhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7146615/pdf
dc.identifier.unpaywallURLhttps://www.mdpi.com/1424-8220/20/6/1568/pdf?version=1584014289
dc.identifier.urihttp://hdl.handle.net/10668/15236
dc.issue.number6
dc.journal.titleSensors (Basel, Switzerland)
dc.journal.titleabbreviationSensors (Basel)
dc.language.isoen
dc.organizationIBS
dc.pubmedtypeJournal Article
dc.rightsAttribution 4.0 International
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/
dc.subjectdose rate
dc.subjectdosimetry
dc.subjectlight-dependent resistor
dc.subjectradiotherapy
dc.subjectthermal characterization
dc.subject.meshElectrical Equipment and Supplies
dc.subject.meshElectronics, Medical
dc.subject.meshEquipment Design
dc.subject.meshLight
dc.subject.meshModels, Biological
dc.subject.meshRadiometry
dc.subject.meshRadiotherapy Dosage
dc.subject.meshSensitivity and Specificity
dc.subject.meshTemperature
dc.titleLight-Dependent Resistors as Dosimetric Sensors in Radiotherapy.
dc.typeresearch article
dc.type.hasVersionVoR
dc.volume.number20
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PMC7146615.pdf
Size:
3.97 MB
Format:
Adobe Portable Document Format