Publication:
Infective Endocarditis Following Transcatheter Aortic Valve Replacement: Comparison of Balloon- Versus Self-Expandable Valves.

dc.contributor.authorRegueiro, Ander
dc.contributor.authorLinke, Axel
dc.contributor.authorLatib, Azeem
dc.contributor.authorIhlemann, Nikolaj
dc.contributor.authorUrena, Marina
dc.contributor.authorWalther, Thomas
dc.contributor.authorHusser, Oliver
dc.contributor.authorHerrmann, Howard
dc.contributor.authorNombela-Franco, Luis
dc.contributor.authorCheema, Asim
dc.contributor.authorLe Breton, Hervé
dc.contributor.authorStortecky, Stefan
dc.contributor.authorKapadia, Samir
dc.contributor.authorBartorelli, Antonio
dc.contributor.authorSinning, Jan Malte
dc.contributor.authorAmat-Santos, Ignacio
dc.contributor.authorMunoz-Garcia, Antonio
dc.contributor.authorLerakis, Stamatios
dc.contributor.authorGutíerrez-Ibanes, Enrique
dc.contributor.authorAbdel-Wahab, Mohamed
dc.contributor.authorTchetche, Didier
dc.contributor.authorTesta, Luca
dc.contributor.authorEltchaninoff, Helene
dc.contributor.authorLivi, Ugolino
dc.contributor.authorCastillo, Juan Carlos
dc.contributor.authorJilaihawi, Hasan
dc.contributor.authorWebb, John
dc.contributor.authorBarbanti, Marco
dc.contributor.authorKodali, Susheel
dc.contributor.authorde Brito Jr, Fabio
dc.contributor.authorRibeiro, Henrique
dc.contributor.authorMiceli, Antonio
dc.contributor.authorFiorina, Claudia
dc.contributor.authorActis Dato, Guglielmo Mario
dc.contributor.authorRosato, Francesco
dc.contributor.authorSerra, Vicenç
dc.contributor.authorMasson, Jean-Bernard
dc.contributor.authorWijeysundera, Harindra
dc.contributor.authorMangione, Jose
dc.contributor.authorFerreira, Maria-Cristina
dc.contributor.authorLima, Valter
dc.contributor.authorCarvalho, Luis
dc.contributor.authorAbizaid, Alexandre
dc.contributor.authorMarino, Marcos
dc.contributor.authorEsteves, Vinicius
dc.contributor.authorAndrea, Julio
dc.contributor.authorMessika-Zeitoun, David
dc.contributor.authorHimbert, Dominique
dc.contributor.authorKim, Won-Keun
dc.contributor.authorPellegrini, Costanza
dc.contributor.authorAuffret, Vincent
dc.contributor.authorNietlispach, Fabian
dc.contributor.authorPilgrim, Thomas
dc.contributor.authorDurand, Eric
dc.contributor.authorLisko, John
dc.contributor.authorMakkar, Raj
dc.contributor.authorLemos, Pedro
dc.contributor.authorLeon, Martin
dc.contributor.authorPuri, Rishi
dc.contributor.authorSan Roman, Alberto
dc.contributor.authorVahanian, Alec
dc.contributor.authorSøndergaard, Lars
dc.contributor.authorMangner, Norman
dc.contributor.authorRodés-Cabau, Josep
dc.date.accessioned2023-02-08T14:37:12Z
dc.date.available2023-02-08T14:37:12Z
dc.date.issued2019-11-07
dc.description.abstractNo data exist about the characteristics of infective endocarditis (IE) post-transcatheter aortic valve replacement (TAVR) according to transcatheter valve type. We aimed to determine the incidence, clinical characteristics, and outcomes of patients with IE post-TAVR treated with balloon-expandable valve (BEV) versus self-expanding valve (SEV) systems. Data from the multicenter Infectious Endocarditis After TAVR International Registry was used to compare IE patients with BEV versus SEV. A total of 245 patients with IE post-TAVR were included (SEV, 47%; BEV, 53%). The timing between TAVR and IE was similar between groups (SEV, 5.5 [1.2-15] months versus BEV, 5.3 [1.7-11.4] months; P=0.89). Enterococcal IE was more frequent in the SEV group (36.5% versus 15.4%; P The characteristics of IE post-TAVR, including microorganism type, vegetation location, and embolic complications but not early or late mortality, differed according to valve type. These results may help to guide the diagnosis and management of IE and inform future research studies in the field.
dc.identifier.doi10.1161/CIRCINTERVENTIONS.119.007938
dc.identifier.essn1941-7632
dc.identifier.pmid31694412
dc.identifier.urihttp://hdl.handle.net/10668/14644
dc.issue.number11
dc.journal.titleCirculation. Cardiovascular interventions
dc.journal.titleabbreviationCirc Cardiovasc Interv
dc.language.isoen
dc.organizationHospital Universitario Reina Sofía
dc.organizationHospital Universitario Virgen de la Victoria
dc.page.numbere007938
dc.pubmedtypeComparative Study
dc.pubmedtypeJournal Article
dc.pubmedtypeMulticenter Study
dc.subjectendocarditis
dc.subjectincidence
dc.subjectregistry
dc.subjecttranscatheter aortic valve replacement
dc.subject.meshAged
dc.subject.meshAged, 80 and over
dc.subject.meshAortic Valve
dc.subject.meshAortic Valve Stenosis
dc.subject.meshBalloon Valvuloplasty
dc.subject.meshClinical Decision-Making
dc.subject.meshDevice Removal
dc.subject.meshEndocarditis, Bacterial
dc.subject.meshFemale
dc.subject.meshHeart Valve Prosthesis
dc.subject.meshHumans
dc.subject.meshIncidence
dc.subject.meshMale
dc.subject.meshPatient Selection
dc.subject.meshProsthesis Design
dc.subject.meshProsthesis-Related Infections
dc.subject.meshRegistries
dc.subject.meshReoperation
dc.subject.meshRetrospective Studies
dc.subject.meshRisk Assessment
dc.subject.meshRisk Factors
dc.subject.meshTime Factors
dc.subject.meshTranscatheter Aortic Valve Replacement
dc.subject.meshTreatment Outcome
dc.titleInfective Endocarditis Following Transcatheter Aortic Valve Replacement: Comparison of Balloon- Versus Self-Expandable Valves.
dc.typeresearch article
dc.volume.number12
dspace.entity.typePublication

Files