dc.description.abstractLipemia, a significant source of analytical errors in clinical laboratory settings, should be removed prior to measuring biochemical parameters. We investigated whether lipemia in serum/plasma samples can be removed using a method that is easier and more practicable than ultracentrifugation, the current reference method. Seven hospital laboratories in Spain participated in this study. We first compared the effectiveness of ultracentrifugation (108,200×g) and high-speed centrifugation (10,000×g for 15 minutes) in removing lipemia. Second, we compared high-speed centrifugation with two liquid-liquid extraction methods-LipoClear (StatSpin, Norwood, USA), and 1,1,2-trichlorotrifluoroethane (Merck, Darmstadt, Germany). We assessed 14 biochemical parameters: serum/plasma concentrations of sodium ion, potassium ion, chloride ion, glucose, total protein, albumin, creatinine, urea, alkaline phosphatase, gamma-glutamyl transferase, alanine aminotransferase, aspartate-aminotransferase, calcium, and bilirubin. We analyzed whether the differences between lipemia removal methods exceeded the limit for clinically significant interference (LCSI). When ultracentrifugation and high-speed centrifugation were compared, no parameter had a difference that exceeded the LCSI. When high-speed centrifugation was compared with the two liquid-liquid extraction methods, we found differences exceeding the LCSI in protein, calcium, and aspartate aminotransferase in the comparison with 1,1,2-trichlorotrifluoroethane, and in protein, albumin, and calcium in the comparison with LipoClear. Differences in other parameters did not exceed the LCSI. High-speed centrifugation (10,000×g for 15 minutes) can be used instead of ultracentrifugation to remove lipemia in serum/plasma samples. LipoClear and 1,1,2-trichlorotrifluoroethane are unsuitable as they interfere with the measurement of certain parameters.
dc.identifier.doi10.3343/alm.2018.38.6.518
dc.identifier.essn2234-3814
dc.identifier.pmcPMC6056396
dc.identifier.pmid30027694
dc.identifier.pubmedURLhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC6056396/pdf
dc.identifier.unpaywallURLhttp://www.annlabmed.org/journal/download_pdf.php?doi=10.3343/alm.2018.38.6.518
dc.identifier.urihttp://hdl.handle.net/10668/12730
dc.issue.number6
dc.journal.titleAnnals of laboratory medicine
dc.journal.titleabbreviationAnn Lab Med
dc.language.isoen
dc.organizationHospital Universitario Virgen del Rocío
dc.page.number518-523
dc.pubmedtypeComparative Study
dc.pubmedtypeJournal Article
dc.pubmedtypeMulticenter Study
dc.rightsAttribution-NonCommercial-ShareAlike 4.0 International
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/4.0/
dc.subject1,1,2-trichlorotrifluoroethane
dc.subjectHigh-speed centrifugation
dc.subjectInterference
dc.subjectLipemia
dc.subjectLipid removal method
dc.subjectLipoClear
dc.subject.meshAlanine Transaminase
dc.subject.meshCalcium
dc.subject.meshCentrifugation
dc.subject.meshCreatinine
dc.subject.meshHumans
dc.subject.meshHyperlipidemias
dc.subject.meshLaboratories, Hospital
dc.subject.meshLipids
dc.subject.meshLiquid-Liquid Extraction
dc.titleRemoving Lipemia in Serum/Plasma Samples: A Multicenter Study.
dc.typeresearch article
dc.type.hasVersionVoR
dc.volume.number38
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PMC6056396.pdf
Size:
219.15 KB
Format:
Adobe Portable Document Format