Publication: Removing Lipemia in Serum/Plasma Samples: A Multicenter Study.
dc.contributor.author | Castro-Castro, María José | |
dc.contributor.author | Candás-Estébanez, Beatriz | |
dc.contributor.author | Esteban-Salán, Margarita | |
dc.contributor.author | Calmarza, Pilar | |
dc.contributor.author | Arrobas-Velilla, Teresa | |
dc.contributor.author | Romero-Román, Carlos | |
dc.contributor.author | Pocoví-Mieras, Miguel | |
dc.contributor.author | Aguilar-Doreste, José Ángel | |
dc.contributor.author | Commission on Lipoprotein and Vascular Diseases, Sociedad Española de Química Clínica | |
dc.date.accessioned | 2023-01-25T10:20:59Z | |
dc.date.available | 2023-01-25T10:20:59Z | |
dc.date.issued | 2018 | |
dc.description.abstract | Lipemia, a significant source of analytical errors in clinical laboratory settings, should be removed prior to measuring biochemical parameters. We investigated whether lipemia in serum/plasma samples can be removed using a method that is easier and more practicable than ultracentrifugation, the current reference method. Seven hospital laboratories in Spain participated in this study. We first compared the effectiveness of ultracentrifugation (108,200×g) and high-speed centrifugation (10,000×g for 15 minutes) in removing lipemia. Second, we compared high-speed centrifugation with two liquid-liquid extraction methods-LipoClear (StatSpin, Norwood, USA), and 1,1,2-trichlorotrifluoroethane (Merck, Darmstadt, Germany). We assessed 14 biochemical parameters: serum/plasma concentrations of sodium ion, potassium ion, chloride ion, glucose, total protein, albumin, creatinine, urea, alkaline phosphatase, gamma-glutamyl transferase, alanine aminotransferase, aspartate-aminotransferase, calcium, and bilirubin. We analyzed whether the differences between lipemia removal methods exceeded the limit for clinically significant interference (LCSI). When ultracentrifugation and high-speed centrifugation were compared, no parameter had a difference that exceeded the LCSI. When high-speed centrifugation was compared with the two liquid-liquid extraction methods, we found differences exceeding the LCSI in protein, calcium, and aspartate aminotransferase in the comparison with 1,1,2-trichlorotrifluoroethane, and in protein, albumin, and calcium in the comparison with LipoClear. Differences in other parameters did not exceed the LCSI. High-speed centrifugation (10,000×g for 15 minutes) can be used instead of ultracentrifugation to remove lipemia in serum/plasma samples. LipoClear and 1,1,2-trichlorotrifluoroethane are unsuitable as they interfere with the measurement of certain parameters. | |
dc.identifier.doi | 10.3343/alm.2018.38.6.518 | |
dc.identifier.essn | 2234-3814 | |
dc.identifier.pmc | PMC6056396 | |
dc.identifier.pmid | 30027694 | |
dc.identifier.pubmedURL | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6056396/pdf | |
dc.identifier.unpaywallURL | http://www.annlabmed.org/journal/download_pdf.php?doi=10.3343/alm.2018.38.6.518 | |
dc.identifier.uri | http://hdl.handle.net/10668/12730 | |
dc.issue.number | 6 | |
dc.journal.title | Annals of laboratory medicine | |
dc.journal.titleabbreviation | Ann Lab Med | |
dc.language.iso | en | |
dc.organization | Hospital Universitario Virgen del Rocío | |
dc.page.number | 518-523 | |
dc.pubmedtype | Comparative Study | |
dc.pubmedtype | Journal Article | |
dc.pubmedtype | Multicenter Study | |
dc.rights | Attribution-NonCommercial-ShareAlike 4.0 International | |
dc.rights.accessRights | open access | |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-sa/4.0/ | |
dc.subject | 1,1,2-trichlorotrifluoroethane | |
dc.subject | High-speed centrifugation | |
dc.subject | Interference | |
dc.subject | Lipemia | |
dc.subject | Lipid removal method | |
dc.subject | LipoClear | |
dc.subject.mesh | Alanine Transaminase | |
dc.subject.mesh | Calcium | |
dc.subject.mesh | Centrifugation | |
dc.subject.mesh | Creatinine | |
dc.subject.mesh | Humans | |
dc.subject.mesh | Hyperlipidemias | |
dc.subject.mesh | Laboratories, Hospital | |
dc.subject.mesh | Lipids | |
dc.subject.mesh | Liquid-Liquid Extraction | |
dc.title | Removing Lipemia in Serum/Plasma Samples: A Multicenter Study. | |