Publication:
Modified Geometry of 106Ru Asymmetric Eye Plaques to Improve Dosimetric Calculations in Ophthalmic Brachytherapy.

dc.contributor.authorMiras, Héctor
dc.contributor.authorTerrón, José Antonio
dc.contributor.authorBertolet, Alejandro
dc.contributor.authorLeal, Antonio
dc.date.accessioned2023-05-03T14:11:40Z
dc.date.available2023-05-03T14:11:40Z
dc.date.issued2022-04-29
dc.description.abstractRu/Rh asymmetric plaques for ophthalmic brachytherapy have special geometric designs with a cutout intended to prevent irradiation of critical ocular structures proximal to the tumor. In this work, we present new geometric models for PENELOPE+PenEasy Monte Carlo simulations of these applicators, differing from the vendor-reported geometry, that better match their real geometry to assess their dosimetric impact. Simulation results were benchmarked to experimental dosimetric data from radiochromic film measurements, data provided by the manufacturer in the calibration certificates, and other experimental results published in the literature, obtaining, in all cases, better agreement with the modified geometries. The clinical impact of the new geometric models was evaluated by simulating real clinical cases using patient-specific eye models. The cases calculated using the modified geometries presented higher doses to the critical structures proximal to the cutout region. The modified geometric models presented in this work provide a more accurate representation of the asymmetric plaques, greatly improving the agreement between Monte Carlo calculations and experimental measurements. Lack of consideration of accurate geometric models has been shown to be translated into notable increases in dose to organs at risk in clinical cases.
dc.identifier.doi10.3390/jpm12050723
dc.identifier.issn2075-4426
dc.identifier.pmcPMC9144752
dc.identifier.pmid35629144
dc.identifier.pubmedURLhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC9144752/pdf
dc.identifier.unpaywallURLhttps://www.mdpi.com/2075-4426/12/5/723/pdf?version=1652233712
dc.identifier.urihttp://hdl.handle.net/10668/21377
dc.issue.number5
dc.journal.titleJournal of personalized medicine
dc.journal.titleabbreviationJ Pers Med
dc.language.isoen
dc.organizationHospital Universitario Virgen Macarena
dc.organizationHospital Universitario Virgen Macarena
dc.organizationInstituto de Biomedicina de Sevilla-IBIS
dc.pubmedtypeJournal Article
dc.rightsAttribution 4.0 International
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/
dc.subjectMonte Carlo
dc.subjectophthalmic brachytherapy
dc.subjectruthenium plaque
dc.titleModified Geometry of 106Ru Asymmetric Eye Plaques to Improve Dosimetric Calculations in Ophthalmic Brachytherapy.
dc.typeresearch article
dc.type.hasVersionVoR
dc.volume.number12
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PMC9144752.pdf
Size:
5.07 MB
Format:
Adobe Portable Document Format