Publication:
Air density dependence of the response of the PTW SourceCheck 4pi ionization chamber for 125I brachytherapy seeds.

dc.contributor.authorTorres Del Río, J
dc.contributor.authorTornero-López, A M
dc.contributor.authorGuirado, D
dc.contributor.authorPérez-Calatayud, J
dc.contributor.authorLallena, A M
dc.date.accessioned2023-01-25T09:47:28Z
dc.date.available2023-01-25T09:47:28Z
dc.date.issued2017-05-23
dc.description.abstractTo analyze the air density dependence of the response of the new SourceCheck 4pi ionization chamber, manufactured by PTW. The air density dependence of three different SourceCheck 4pi chambers was studied by measuring 125I sources. Measurements were taken by varying the pressure from 746.6 to 986.6hPa in a pressure chamber. Three different HDR 1000 Plus ionization chambers were also analyzed under similar conditions. A linear and a potential-like function of the air density were fitted to experimental data and their achievement in describing them was analyzed. SourceCheck 4pi chamber response showed a residual dependence on the air density once the standard pressure and temperature factor was applied. The chamber response was overestimated when the air density was below that under normal atmospheric conditions. A similar dependence was found for the HDR 1000 Plus chambers analyzed. A linear function of the air density permitted a very good description of this residual dependence, better than with a potential function. No significant variability between the different specimens of the same chamber model studied was found. The effect of overestimation observed in the chamber responses once they are corrected for the standard pressure and temperature may represent a non-negligible ∼4% overestimation in high altitude cities as ours (700m AMSL). This overestimation behaves linearly with the air density in all cases analyzed.
dc.identifier.doi10.1016/j.ejmp.2017.05.056
dc.identifier.essn1724-191X
dc.identifier.pmid28610703
dc.identifier.unpaywallURLhttp://arxiv.org/pdf/1809.09046
dc.identifier.urihttp://hdl.handle.net/10668/11300
dc.journal.titlePhysica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
dc.journal.titleabbreviationPhys Med
dc.language.isoen
dc.organizationHospital Universitario San Cecilio
dc.page.number93-97
dc.pubmedtypeJournal Article
dc.rights.accessRightsopen access
dc.subjectAir density dependence
dc.subjectBrachytherapy
dc.subjectIonization chamber
dc.subjectLow-energy sources
dc.subjectPTW SourceCheck 4pi
dc.subject.meshAir
dc.subject.meshBrachytherapy
dc.subject.meshIodine Radioisotopes
dc.subject.meshPressure
dc.subject.meshRadiometry
dc.subject.meshTemperature
dc.titleAir density dependence of the response of the PTW SourceCheck 4pi ionization chamber for 125I brachytherapy seeds.
dc.typeresearch article
dc.type.hasVersionSMUR
dc.volume.number38
dspace.entity.typePublication

Files