Publication: Structural Damage of Jailed Guidewire During the Treatment of Coronary Bifurcation Lesions: A Microscopic Randomized Trial.
Loading...
Identifiers
Date
2016-06-15
Authors
Pan, Manuel
Ojeda, Soledad
Villanueva, Elena
Chavarria, Jorge
Romero, Miguel
Suarez de Lezo, Javier
Mazuelos, Francisco
Segura, Jose
Carrasco, Francisco
Hidalgo, Francisco
Advisors
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Abstract
The study sought to compare the safety (resistance to damage) and efficacy (ability to cross the side branch) of polymer-coated and non-polymer-coated guidewires in the jailed wire technique used during the percutaneous treatment of bifurcation lesions. The jailed wire technique is a useful strategy in the treatment of bifurcation lesions by provisional stenting. However, these wires can be damaged or even be broken during their removal. We performed a randomized study in patients with bifurcation lesions treated by provisional stenting. The jailed wire technique was mandatory, and the types of guidewires, polymer-coated (n = 115) and non-polymer-coated (n = 120), were randomized. After the procedures, the wires were evaluated by stereoscopic microscopy. The induced damage in the wires was classified as follows: no damage, mild, moderate, or severe. The clinical characteristics were similar between patients treated with polymer-coated or non-polymer-coated wires. Polymer-coated wires were significantly (p Jailed wires during interventional procedures of bifurcation lesions commonly showed microscopic damage. Polymer-coated wires were more resistant to retrieval damage and were more efficient in crossing the side branch ostium than non-polymer-coated wires. (Jailed Wire Technique in the Treatment of Coronary Bifurcations Lesions With Stent: Stereoscopic Microscopy Study; NCT02516891).
Description
MeSH Terms
Aged
Cardiac catheters
Coated materials, biocompatible
Coronary angiography
Coronary artery disease
Equipment design
Equipment failure
Female
Humans
Male
Microscopy
Middle aged
Percutaneous coronary intervention
Polymers
Prospective studies
Single-blind method
Spain
Stents
Treatment outcome
Cardiac catheters
Coated materials, biocompatible
Coronary angiography
Coronary artery disease
Equipment design
Equipment failure
Female
Humans
Male
Microscopy
Middle aged
Percutaneous coronary intervention
Polymers
Prospective studies
Single-blind method
Spain
Stents
Treatment outcome
DeCS Terms
Angiografía coronaria
Catéteres cardíacos
Diseño de equipo
Enfermedad de la arteria coronaria
Falla de equipo
Intervención coronaria percutánea
Materiales biocompatibles revestidos
Microscopía
Catéteres cardíacos
Diseño de equipo
Enfermedad de la arteria coronaria
Falla de equipo
Intervención coronaria percutánea
Materiales biocompatibles revestidos
Microscopía
CIE Terms
Keywords
Bifurcation lesions, Jailed wire technique, Provisional stenting
Citation
Pan M, Ojeda S, Villanueva E, Chavarria J, Romero M, Suarez de Lezo J, et al. Structural Damage of Jailed Guidewire During the Treatment of Coronary Bifurcation Lesions: A Microscopic Randomized Trial. JACC Cardiovasc Interv. 2016 Sep 26;9(18):1917-24