Publication: Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr.
dc.contributor.author | Calvo, Jorge | |
dc.contributor.author | Cano, María Eliecer | |
dc.contributor.author | Pitart, Cristina | |
dc.contributor.author | Marco, Francesc | |
dc.contributor.author | Rodríguez-Martínez, José Manuel | |
dc.contributor.author | Pascual, Alvaro | |
dc.contributor.author | Martínez-Martínez, Luis | |
dc.contributor.authoraffiliation | [Calvo,J; Cano,ME; Martínez-Martínez,L] Service of Microbiology, University Hospital Marqués de Valdecilla-IFIMAV, Santander. [Pitart,C; Marco,F] Service of Microbiology, Hospital Clínic, Barcelona. [Rodríguez-Martínez,JM; Pascual,A] Department of Microbiology, School of Medicine, University of Seville. University Hospital Virgen Macarena, Seville. [Martínez-Martínez,L] Department of Molecular Biology, School of Medicine, University of Cantabria, Santander, Spain. | es |
dc.contributor.funder | This work was supported by the Consejería de Innovación Ciencia y Empresa, Junta de Andalucía (P07-CTS-02908), Spain. It was partly supported by the Ministerio de Sanidad y Consumo, Instituto de Salud Carlos III-FEDER (grant PI050690), the Spanish Network for Research in Infectious Diseases (REIPI RD06/0008), and the Consejería de Salud, Junta de Andalucía (PI-0282-2010). | |
dc.date.accessioned | 2014-04-14T12:11:08Z | |
dc.date.available | 2014-04-14T12:11:08Z | |
dc.date.issued | 2011-09 | |
dc.description | Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't; | es |
dc.description.abstract | The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted. | es |
dc.description.version | Yes | es |
dc.identifier.citation | Calvo J, Cano ME, Pitart C, Marco F, Rodríguez-Martínez JM, Pascual A, et al. Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr. J Clin Microbiol. 2011; 49(9):3343-5 | es |
dc.identifier.doi | 10.1128/JCM.00563-11 | |
dc.identifier.essn | 1098-660X | |
dc.identifier.issn | 0095-1137 | |
dc.identifier.pmc | PMC3165611 | |
dc.identifier.pmid | 21775549 | |
dc.identifier.uri | http://hdl.handle.net/10668/1580 | |
dc.journal.title | Journal of clinical microbiology | |
dc.language.iso | en | |
dc.publisher | American Society for Microbiology | es |
dc.relation.publisherversion | http://jcm.asm.org/content/49/9/3343.abstract | es |
dc.rights.accessRights | restricted access | |
dc.subject | Escherichia coli | es |
dc.subject | Aminoglicósidos | es |
dc.subject | Enterobacteriaceae | es |
dc.subject | Klebsiella pneumoniae | es |
dc.subject | Errores diagnósticos | es |
dc.subject | Antibacterianos | es |
dc.subject.mesh | Medical Subject Headings::Chemicals and Drugs::Chemical Actions and Uses::Pharmacologic Actions::Therapeutic Uses::Anti-Infective Agents::Anti-Bacterial Agents | es |
dc.subject.mesh | Medical Subject Headings::Technology, Industry, Agriculture::Technology, Industry, and Agriculture::Technology::Automation | es |
dc.subject.mesh | Medical Subject Headings::Analytical, Diagnostic and Therapeutic Techniques and Equipment::Diagnosis::Diagnostic Errors | es |
dc.subject.mesh | Medical Subject Headings::Organisms::Bacteria::Gram-Negative Bacteria::Gram-Negative Facultatively Anaerobic Rods::Enterobacteriaceae | es |
dc.subject.mesh | Medical Subject Headings::Phenomena and Processes::Genetic Phenomena::Genetic Structures::Genome::Genome Components::Genes::Genes, Bacterial | es |
dc.subject.mesh | Medical Subject Headings::Organisms::Eukaryota::Animals::Chordata::Vertebrates::Mammals::Primates::Haplorhini::Catarrhini::Hominidae::Humans | es |
dc.subject.mesh | Medical Subject Headings::Analytical, Diagnostic and Therapeutic Techniques and Equipment::Diagnosis::Diagnostic Techniques and Procedures::Clinical Laboratory Techniques::Microbiological Techniques::Microbial Sensitivity Tests | es |
dc.subject.mesh | Medical Subject Headings::Chemicals and Drugs::Heterocyclic Compounds::Heterocyclic Compounds, 2-Ring::Quinolines::Quinolones | es |
dc.subject.mesh | Medical Subject Headings::Chemicals and Drugs::Carbohydrates::Glycosides::Aminoglycosides | es |
dc.title | Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr. | es |
dc.type | research article | |
dc.type.hasVersion | VoR | |
dspace.entity.type | Publication |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- CalvoJ_EvaluationOfThreeAutomated.pdf
- Size:
- 124 KB
- Format:
- Adobe Portable Document Format
- Description:
- Artículo publicado