Publication:
Use and Misuse of Cq in qPCR Data Analysis and Reporting

dc.contributor.authorRuiz-Villalba, Adrián
dc.contributor.authorRuijter, Jan M.
dc.contributor.authorvan den Hoff, Maurice J. B.
dc.contributor.authoraffiliation[Ruiz-Villalba,A] Department of Animal Biology, Faculty of Sciences, Instituto Malagueño de Biomedicina (IBIMA), University of Málaga, Málaga, Spain. [Ruiz-Villalba,A] BIONAND, Centro Andaluz de Nanomedicina y Biotecnología, Junta de Andalucía, Universidad de Málaga, Málaga, Spain. [Ruijter,JM; van den Hoff, M.J.B] Department of Medical Biology, Amsterdam University Medical Centres, Location Academic Medical Center, Meibergdreef 15, Amsterdam, The Netherlands.
dc.contributor.funderA.R.V. is supported by funds from University of Málaga (Incorporación de doctores from the I Plan Propio de Incorporación de Doctores, 2020).
dc.date.accessioned2022-04-01T13:01:40Z
dc.date.available2022-04-01T13:01:40Z
dc.date.issued2021-05-29
dc.description.abstractIn the analysis of quantitative PCR (qPCR) data, the quantification cycle (Cq) indicates the position of the amplification curve with respect to the cycle axis. Because Cq is directly related to the starting concentration of the target, and the difference in Cq values is related to the starting concentration ratio, the only results of qPCR analysis reported are often Cq, ΔCq or ΔΔCq values. However, reporting of Cq values ignores the fact that Cq values may differ between runs and machines, and, therefore, cannot be compared between laboratories. Moreover, Cq values are highly dependent on the PCR efficiency, which differs between assays and may differ between samples. Interpreting reported Cq values, assuming a 100% efficient PCR, may lead to assumed gene expression ratios that are 100-fold off. This review describes how differences in quantification threshold setting, PCR efficiency, starting material, PCR artefacts, pipetting errors and sampling variation are at the origin of differences and variability in Cq values and discusses the limits to the interpretation of observed Cq values. These issues can be avoided by calculating efficiency-corrected starting concentrations per reaction. The reporting of gene expression ratios and fold difference between treatments can then easily be based on these starting concentrations.es_ES
dc.description.versionYeses_ES
dc.identifier.citationRuiz-Villalba A, Ruijter JM, van den Hoff MJB. Use and Misuse of Cq in qPCR Data Analysis and Reporting. Life. 2021 May 29;11(6):496es_ES
dc.identifier.doi10.3390/life11060496es_ES
dc.identifier.essn2075-1729
dc.identifier.pmcPMC8229287
dc.identifier.pmid34072308es_ES
dc.identifier.urihttp://hdl.handle.net/10668/3514
dc.journal.titleLife
dc.language.isoen
dc.page.number22 p.
dc.publisherMDPIes_ES
dc.relation.publisherversionhttps://www.mdpi.com/2075-1729/11/6/496es_ES
dc.rightsAtribución 4.0 Internacional*
dc.rights.accessRightsAcceso abiertoes_ES
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.subjectqPCR analysises_ES
dc.subjectCqes_ES
dc.subjectQuantification cyclees_ES
dc.subjectQuantification thresholdes_ES
dc.subjectPCR efficiencyes_ES
dc.subjectPoisson variationes_ES
dc.subjectLODes_ES
dc.subjectLOQes_ES
dc.subjectArtefactoses_ES
dc.subjectLaboratorioses_ES
dc.subjectExpresión génicaes_ES
dc.subjectReacción en cadena en tiempo real de la polimerasaes_ES
dc.subject.meshMedical Subject Headings::Analytical, Diagnostic and Therapeutic Techniques and Equipment::Investigative Techniques::Artifactses_ES
dc.subject.meshMedical Subject Headings::Health Care::Health Care Facilities, Manpower, and Services::Health Facilities::Laboratorieses_ES
dc.subject.meshMedical Subject Headings::Phenomena and Processes::Genetic Phenomena::Genetic Processes::Gene Expressiones_ES
dc.subject.meshMedical Subject Headings::Analytical, Diagnostic and Therapeutic Techniques and Equipment::Investigative Techniques::Genetic Techniques::Nucleic Acid Amplification Techniques::Polymerase Chain Reaction::Real-Time Polymerase Chain Reactiones_ES
dc.titleUse and Misuse of Cq in qPCR Data Analysis and Reportinges_ES
dc.typereview article
dc.type.hasVersionVoR
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
RuizVillalba_UseAndMisuse.pdf
Size:
3.19 MB
Format:
Adobe Portable Document Format
Description:
Revisión